Mineral Water (2) (SI-90 4E)
Please use this column with suppressor type ion chromatography system.
Mineral water for drinking was analyzed using IC
SI-90 4E (a column for anions analysis).

Sample :
Mineral water(
Water), 20micro-L
1.
F-
2.
Cl-
3.
NO3-
4.
HCO3-
5.
SO42-
Column : Shodex IC SI-90 4E (4.0mmID*250mm)
Eluent : 1.8mM Na2CO3 + 1.7mM NaHCO3 aq.
Flow rate : 1.0mL/min
Detector : Suppressed conductivity
Column temp. : 25deg-C