Anion Standards and Organic Acids (2) (SI-90 4E)
Please use this column with suppressor type ion chromatography system.
Anion standards and organic acids were analyzed using IC
SI-90 4E (a column for anions analysis).

Sample : 20micro-L
1.
F- 2mg/L
2.
Acetic acid 10mg/L
3.
Formic acid 2mg/L
4.
Methacrylic acid 10mg/L
5.
Cl- 3mg/L
6.
NO2- 5mg/L
7.
Br- 10mg/L
8.
NO3- 10mg/L
9.
PO43- 15mg/L
10.
SO42- 15mg/L
11.
Oxalic acid 15mg/L
Column : Shodex IC SI-90 4E (4.0mmID*250mm)
Eluent : 12mM NaHCO3 aq.
Flow rate : 1.0mL/min
Detector : Suppressed conductivity