Producto code |
Y4617000
|
Y4617004
|
---|---|---|
Model
|
ERC-3215alpha
|
ERC-3415alpha
|
Solvent channels |
2 channels
|
4 channels
|
Weight |
ca 5.1kg
|
ca 5.5kg
|
Degassing Capacity |
When ion-exchanged water saturated with air at
25deg-C is put through at a flow rate of 3mL/min,
no bubble is observed at the outlet of the apparatus. (Measured dissolved oxygen level at the outlet : 2ppm at flow rate of 3mL/min) |
|
---|---|---|
Internal volume |
7mL/channel
|
|
Max. flow rate |
20mL/min for each channel (Eluent: 25deg-C pure
water)
|
|
External output |
An open connector signal is delivered to the
external output signal terminal,
When "PRES" and "LEAK" LEDs light. |
|
Dimensions |
W71 x D310 x H136 (mm)
|
|
Power source |
AC 100 to 240 V, 50/60Hz
|
Functional Displays |
Power On/Off display
|
"POWER" LED lights, when the power is supplied.
|
---|---|---|
Status monitoring
function |
"READY" LED lights, when the internal pressure
in the vacuum chamber
is below a predetermined limit. |
|
Pressure monitoring function
|
"PRES" LED lights, when the internal pressurein
the vacuum chamber
does not reach a predetermined level within a predetermined time. |
|
Leak monitor
|
"LEAK" LED lights, when the liquid leaks in the
apparatus.
|
|
Self cleaning
|
The vacuum line is cleaned by air suction.
|
|
Vacuum pump operation
switching function |
NORM.: Controlled operation in normal run
CONT.: Cntinuous operation in case og high degree of degassing degassing |