Purine Bases and Pyrimidine Bases (1) (GS-320 HQ)
Purine bases and pyrimidine bases were analyzed using Asahipak
GS-320
HQ (a multimode column).

Sample : 50micro-g/mL each, 20micro-L
1.
Cytosine
2.
Adenine
3.
Guanine
4.
Uracil
5.
Hypoxanthine
6.
Thymine
Column : Shodex Asahipak GS-320 HQ (7.5mmID*300mm)
Eluent : 50mM NaH2PO4 aq./50mM H3PO4 aq.=500/380
Flow rate : 1.0mL/min
Detector : UV(260nm)
Column temp. : 30deg-C