Effect of Sample Injection Volume
Effect of sample injection volume when Asahipak 
NH2P-50 4E 
is used is shown. When making sample solution to be injected, sample should be dissolved in a solvent whose composition is as 
close as the eluent, preferably in the eluent. When 75% acetonitrile aqueous solution is used for the eluent and a sample can be dissolved only in aqueous solution 
(or 50v/v% ethanol, etc.), injection volume should be as small as possible. 

Sample : 20micro-g each in purified water
1. 
Fructose
2. 
Glucose
3. 
Sucrose
4. 
Lactose
5. 
Maltose
Column       : Shodex Asahipak NH2P-50 4E (4.6mmID*250mm)
Eluent       : CH3CN/H2O=75/25
Flow rate    : 1.0mL/min
Detector     : Shodex RI
Column temp. : 25deg-C