Dietary Fiber
HPLC method is used for the quantitative analysis of low molecular weight dietary 
fibers. In this analysis, it is necessary to divide the sample into two fractions, 
one is hard of digestion fraction composed of saccharides equal or larger than 
trisaccharide (hard of digestion oligosaccharides + dietary fibers) and another 
is easy of digestion fraction composed of monosaccharides, disaccharides and sugar 
alcohols. And, the portion of hard of digestion fraction can be considered as 
the portion of low molecular weight dietary fibers. Analysis of the low molecule 
water solubility dietary fiber which used SUGAR 
KS-802 
and Asahipak 
GS-220 HQ was compared. Using GS-220 HQ, monosaccharides, 
disaccharides and sugar alcohols elute after the position indicated by the arrow 
mark and it is easy to divide the two fractions. Therefore, the column can be 
used for the quantitative analysis of low molecular dietary fibers. Using KS-802, 
the separation of monosaccharides and disaccharides is better.

Sample : 
Dietary fiber
Columns      : Shodex Asahipak GS-220 HQ (7.5mmID*300mm) x 2   Shodex SUGAR KS-802 (8.0mmID*300mm) x 2  
Eluent       : H2O                                             H2O
Flow rate    : 0.5mL/min                                       0.3mL/min
Detector     : Shodex RI                                       Shodex RI
Column temp. : 60deg-C                                         80deg-C