Saccharides and Amino Acids (1)
Generally, ligand exchange mode columns are used with aqueous eluent, but the 
addition of the same cation as the counter ion enables the columns to separate 
samples under ion exchange mode additionally to the ligand exchange mode. CaSO
4 
aqueous solution is used for SUGAR 
SC1011 as its counter 
ion is Ca
2+. This means that ligand exchange occurs 
between the counter ions of the packing material and the hydroxyl groups of saccharides 
and ion exchange occurs between the sulfonate groups and the amino groups of the 
saccharides. For sensitive measurement, attention should be paid to the positions 
of minus peak due to the cations in the eluent.

Sample :
1. 
Maltose
2. 
Glucose
3. 
Fructose
4. 
Glutamic acid
5. 
Aspartic acid
6. 
Threonine
7. 
Serine
8. 
Glutamine
9. 
Alanine
Column       : Shodex SUGAR SC1011 (8.0mmID*300mm)
Eluent       : 0.01M CaSO4(pH6.0)
Flow rate    : 1.0mL/min
Detector     : Shodex RI
Column temp. : 80deg-C