Basic Drugs : Effect of Ion Concentration
Effect of ion concentration on elution pattern for the analysis of scopolamine 
which has tertiary amine and whose pKa is 7.6 is shown. When the ion concentration 
is lowered, non-specific adsorption is observed for ODS-A and sample is completely 
adsorbed at the ion concentration lower than 10mM. This tendency is also observed 
even for ODS-C which is a polymer-coated ODS column and sample is completely adsorbed 
at 0mM. To the contrary, Asahipak 
ODP columns are free 
from the non-specific adsorption problem at any ion concentration level. Then 
ODP columns are suitable for analysis with the ESI method (LC/MS) in which ionization 
of sample is often supressed under high-salt concentration.

Sample : 1.0mg/mL 
Scopolamine, 50micro-L
 
Columns      : Shodex Asahipak ODP-50 4D (4.6mmID*150mm), ODS-A, ODS-C (4.6mmID*150mm each)
Eluent       : Phosphate buffer(pH7)/CH3CN=60/40
Flow rate    : 0.6mL/min
Detector     : UV(254nm)
Column temp. : 30deg-C