Rubber (Toluene Eluent)
Natural rubber was analyzed using GPC
KF-802 and
KF-806M (columns for organic SEC (GPC)). Since
natural rubber is easily dissolved in toluene, toluene eluent is used for
GPC analysis of this sample.

Sample :
Rubber
Columns : Shodex GPC KF-806M x 2 + KF-802 (8.0mmID*300mm each)
Eluent : Toluene
Flow rate : 1.0mL/min
Detector : Shodex RI
Column temp. : Room temp.