Sulfate and Sulfite
This chromatogram shows a data of analysis of standard anions using IC
I-524A with an alkaline
eluent. When the pH of the eluent is higher than 7.0, 10% methanol should be added to maintain the durability of the packing material.

Sample : 10mg/L each, 100micro-L
1.
SO42-
2.
SO32-
3.
S2O32-
4.
SCN-
Column : Shodex IC I-524A (4.6mmID*100mm)
Eluent : 0.25mM p-Hydroxybenzoic acid + 1.2mM N,N-Diethylethanolamine + 10% CH3OH
Flow rate : 1.5mL/min
Detector : Non-suppressed coductivity
Column temp. : 50deg-C