Sea Water (1) (I-524A) 
Anions in sea water were analyzed using IC 
I-524A. Since sea water contains a considerable amount of chloride ion, 
the chlorine should be removed prior to analysis. Put Ag-type cation exchanger in a disposable syringe and pass the sea water through 
the syringe. After the pretreatment, other anions can be analyzed without the interference from chloride ion.

Sample : 
Sea water (
Water)
1.
F-
2.
SO42-
Column       : Shodex IC I-524A (4.6mmID*100mm)
Eluent       : 2.5mM Phthalic acid(pH4.0 adjusted with Tris(hydroxymethyl)aminomethane)
Flow rate    : 1.2mL/min
Detector     : Non-suppressed coductivity
Column temp. : 40deg-C