Effect of Temperature (YS-50)
The effect of temperature for IC
YS-50 was investigated. This shows no significant influence of temperature on elution.


Sample : 10micro-L
1.
Li+ 2mg/L
2.
Na+ 10mg/L
3.
NH4+ 10mg/L
4.
K+ 20mg/L
5.
Mg2+ 10mg/L
6.
Ca2+ 20mg/L
Column : Shodex IC YS-50 (4.6mmID*125mm)
Eluent : 4mM Methanesulfonic acid aq.
Flow rate : 1.0mL/min
Detector : Non-suppressed conductivity