Transition Metal Ions (3) (YS-50)
With 6mM Tartaric asid and 4mM oxalic asid aqueous solution as the eluent, transision metals, such as nickel, zinc, and cobalt can be analyzed using IC 
YS-50. 

Sample : 20micro-L
  1. 
Zn2+ 10mg/L
 
  2. 
Co2+ 10mg/L
  3. 
Fe2+ 10mg/L
  4. 
Mn2+ 10mg/L
 
  5. 
Cd2+ 10mg/L
  6. 
Pb2+ 30mg/L
Column       : Shodex IC YS-50 (4.6mmID*125mm) 
Eluent       : 6mM Tartaric acid + 4mM Oxalic acid aq.
Flow rate    : 1.0mL/min
Detector     : Non-suppressed conductivity
Column temp. : 40deg-C