Anion Standards (15) (Comparison of NI-424 and I-524A)
The seven main anions can be separated using a newly developed column, IC 
NI-424 quickly. 
The theoretical plate number of the column is more than double that of the existing column IC 
I-524A.
Comparison with I-524A column:
1) NI-424 column has achieved separation of phosphoric acid ion and fluoride ion 
which was difficult to sepatate using I-524A.
2) With NI-424, each peak is sharper, and separation of ions and peak balance 
have been improved. In particular, improvement has been made in the separation 
of chloride and nitrite ions making NI-424 column suited to the analysis of a 
trace amount of nitrite ion contained in high-concentration chloride ion; demand 
for such analysis is high in the area of environmental analysis.

Sample : 20micro-L
1. 
H2PO4- 10mg/L
2. 
F- 1mg/L
3. 
Cl- 1mg/L
4. 
NO2- 5mg/L
5. 
Br- 5mg/L
6. 
NO3- 5mg/L
7. 
SO42- 5mg/L
(Left)
Column       : Shodex IC I-524A (4.6mmID*100mm)
Eluent       : 2.5mM Phthalic acid + 2.3mM Tris(hydroxymethyl)aminomethane aq.
Flow rate    : 1.2mL/min
Detector     : Non-suppressed coductivity
Column temp. : 40deg-C
(Right)
Column       : Shodex IC NI-424 (4.6mmID*100mm)
Eluent       : 8mM 4-Hydroxybenzoic acid + 2.8mM Bis-Tris + 2mM Phenylboronic acid + 0.005mM *CyDTA aq.
Flow rate    : 1.0mL/min
Detector     : Non-suppressed coductivity
Column temp. : 40deg-C
* CyDTA: trans-1,2-Diaminocyclohexane-N,N,N',N'-tetraacetic acid