Effect of pH on Retention Time (NI-424)
When using IC 
NI-424 and the pH of the eluent is increased on the alkaline side, overall elution shortens. At a pH of 4.5 or higher, 
separation of F
- and  H
2PO
4- is difficult.

Sample : 30micro-L
1. 
H2PO4-
2. 
F-
3. 
Cl-
4. 
NO3-
5. 
SO42-
6. 
I-
7. 
SCN-
Column       : Shodex IC NI-424 (4.6mmID*100mm)
Eluent       : 8mM 4-Hydroxybenzoic acid + 2mM Phenylboronic acid + 0.005mM CyDTA(pH adjusted with Bis-Tris)
Flow rate    : 1.0mL/min
Detector     : Non-suppressed conductivity
Column temp. : 40deg-C