Nucleotides (9) (WA-624 pH4.3)
Nucleotides were analyzed using AXpak
WA-624 (a column
for weak anion exchange chromatography). pH of the eluent is lowered to 4.3. Retention
becomes stronger and separation becomes better. Please compare with the cases of
pH5.0
and
pH5.5
Sample :
1. 5'
CMP
2. 5'
UMP
3. 5'
TMP
4. 5'
AMP
5. 5'
GMP
6. 3'
AMP
7. 3'
GMP
8. 5'
CDP
9. 5'
UTP
Column : Shodex AXpak WA-624 (6.0mmID*150mm)
Eluent : 0.7M CH3COONH4 buffer(pH4.3)
Flow rate : 1.0mL/min
Detector : UV(260nm)
Column temp. : 60deg-C