Saccharides and Amino Sugars (1)
Generally, ligand exchange mode columns are used with aqueous eluent, but
the addition of the same cation as the counter ion enables the columns to
separate samples under ion exchange mode additionally to the ligand
exchange mode. Pb(NO
3)
2 aqueous solution is used as an eluent for SUGAR
SP0810 because the
its counter ion is Pb
2+. This means that ligand exchange occurs between the counter ions of the packing material and
the hydroxyl groups of saccharides and ion exchange occurs between the sulfo groups and the amino groups of the saccharides. Meanwhile, as Pb(NO
3)
2 contains lead, the waste liquid should be disposed
of in accordance with legal procedures.
Sample :
1.
Glucose
2.
Galactose
3.
Mannose
4.
Chloride anion
5.
Glucosamine
6.
Galactosamine
Column : Shodex SUGAR SP0810 (8.0mmID*300mm)
Eluent : 20mM Pb(NO3)2(pH3.8)
Flow rate : 0.7mL/min
Detector : Shodex RI
Column temp. : 60deg-C