Carbonate and Other Anions (1) (I-524A)
Standard anions (F
-, Cl
-,
NO
2-, Br
-, NO
3-,
HPO
42- and SO
42-)
as well as carbonate were analyzed using IC
I-524A. When
an alkaline eluent is used, peaks of divalent cations (Mg
2+,
Ca
2+) may interfere. In such instances, the sample must
be pretreated using an Na
+ type cation exchanger and the divalent cations in the
sample replaced with Na
+.
In case that the pH of the eluent is higher than 7.0, 10% methanol should
be added to maintain the durability of the packing material.
(Preparation of eluent)
1) Put 0.207g of reagent grade p-hydroxybenzoic acid and 100mL of methanol
into a messflask and dissolve.
2) Add pure water to make up a 900mL solution.
3) Add 7.9mL of 0.2M reagent grade diethylethanolamine.
4) Add purified water to make up a 1 liter solution.
5) Pass through a 0.2micron membrane filter before use.
(Remark): 0.2M diethylethanolamine should be kept in a refrigerator.
Sample :
1.
F-
2.
HCO3-
3.
Cl-
4.
NO2-
5.
Br-
6.
NO3-
7.
HPO42-
8.
SO42-
AsO43-
I-
CrO42-
S2O32-
WO42-
MoO42-
SCN-
Column : Shodex IC I-524A (4.6mmID*100mm)
Eluent : 1.5mM p-Hydroxybenzoic acid(pH7.8 adjusted with N,N-Diethylethanolamine) + 10% CH3OH
Flow rate : 1.2mL/min
Detector : Non-suppressed coductivity
Column temp. : 40deg-C