Tap Water (3) (I-524A)
Anions in tap water were analyzed using IC
I-524A. Since
phosphoric ion is not present in this sample, F
-, Cl
-,
NO
3- and SO
42-
can be separated in a short time without t he interference of phosphoric ion peak
using this eluent. When both F
- and H
2PO
4- are present, they cannot be separated each othe with this analysis conditionr.
Sample :
Tap water (
Water), 100micro-L
1.
F-
2.
Cl-
3.
NO3-
4.
SO42-
Column : Shodex IC I-524A (4.6mmID*100mm)
Eluent : 2.5mM Phthalic acid + 2.3mM Tris(hydroxymethyl)aminomethane aq.
Flow rate : 1.2mL/min
Detector : Non-suppressed coductivity
Column temp. : 40deg-C