Serum
The problems in the anaysis using IC
YK-421 is that the
analysis of very small amount of NH
4+ in
a sample which contains large amount of Na
+ is difficult.
Addition of 1.5mM 18-Crown 6-ether in 5mM Tartaric acid + 1mM Dipicolinic acid
+ 1.5g/L Boric acid eluent can solve the problem. Addition of crown 6-ether can
adjust the elution volume of the cations. Especially, the elution volume of K
+
is largely affected by the additon and elute after Mg
2+.
Sample :
Serum
1.
Na+
2.
NH4+
3.
Ca2+
4.
Mg2+
5.
K+
Column : Shodex IC YK-421 (4.6mmID*125mm)
Eluent : (left) 5mM Tartaric acid + 1mM Dipicolinic acid + 1.5g/L Boric acid
(right)5mM Tartaric acid + 1mM Dipicolinic acid + 1.5g/L Boric acid + 1.5mM 18-Crown 6-ether
Flow rate : 1.0mL/min
Detector : Non-suppressed coductivity
Column temp. : 40deg-C