Analysis of Anions in Wine Standard (VH-anion 4D)

Please use this column with suppressor type ion chromatography system.

WINE VH-anion 4D was used to quantify anions in wine standard. VH-anion 4D provides a rapid analysis without any influence from organic acid present in wine. This method using VH-anion 4D for the quantification of anions in wine was co-developed with Institute Heidger (Kesten, Germany). *The wine standard is prepared artificially that contains components that are generally present in wine and commercially available from an institute in Speyer, Germany.


Sample : 50-fold diluted Wine standards, 20micro-L
1. F-
2. Lactic acid
3. Cl-
4. PO43-
5. SO32-
6. SO42-
7. Tartaric acid, Malic acid

Wine Standard
Component Concentration
Original
Sample
Concentration
(1+49)
Diluted Sample
F-
unknown
unknown
Cl-
66.9mg/L
1.338mg/L
PO43-
570.7mg/L
11.414mg/L
SO32-
27.1mg/L
0.542mg/L
SO42-
219.4mg/L
4.388mg/L
Lactic acid
2.062g/L
41.24mg/L
Tartaric acid
2.79g/L
55.8mg/L
Malic acid
3.721g/L
74.42mg/L


Column       : Shodex WINE VH-anionG 4A (4.6mmID*10mm) + WINE VH-anion 4D (4.6mmID*150mm)
Eluent       : 5mM Na2CO3 + 12mM NaHCO3 + 3% Acetone
Flow rate    : 0.8mL/min
Detector     : Suppressed conductivity
Column temp. : 30deg-C