Tap Water (7) (NI-424)

The administrative order of the Ministry of Health, Labour and Welfare #261 (July 22, 2003) specifies the acceptable levels for F-, Cl-, NO2- and NO3-. Their concentrations are measured using ion chromatography. Here, anions in tap water were analyzed using IC NI-424 though NO2- could not be detected in tap water.
Please refer to 'Regulation for Water Supply'.


Sample : Tap water (Water), 100micro-L
1. F-
2. Cl-
3. NO3-
4. SO42-

Column       : Shodex IC NI-424 (4.6mmID*100mm)
Eluent       : 8mM 4-Hydroxybenzoic acid + 2.8mM Bis-Tris + 2mM Phenylboronic acid + 0.005mM *CyDTA aq. 
Flow rate    : 1.0mL/min
Detector     : Non-suppressed coductivity
Column temp. : 40deg-C
* CyDTA: trans-1,2-Diaminocyclohexane-N,N,N',N'-tetraacetic acid