Linearity of Calibration Curves
The figure shows calibration curves of IC
NI-424for three kinds of anions. The calibration curves for chloride and nitrite ions exhibit linearity in a range between 0.01 and 100 mg/L. For phosphoric acid ion which has low sensitivity, the calibration curve exhibits linearity in a range between 0.1 and 100 mg/L.
Sample :
1.
H2PO4-
2.
Cl-
3.
NO2-
Column : Shodex IC NI-424 (4.6mmID*100mm)
Eluent : 2.1mM Phthalic acid + 2.9mM Amino-n-capronic acid + 6mM Phenylboronic acid aq.
Flow rate : 1.0mL/min
Detector : Non-suppressed conductivity
Column temp. : 40deg-C