MW Distribution Analysis (KF-406L HQ)
MW distribution of SRM 706 was measured using organic solvent SEC (GPC) semi-micro column, GPC
KF-406LHQ and conventional organic solvent SEC (GPC) column, GPC
KF-806M. Both columns provided good agreements even using different detector systems (Light scattering and sedimentation equilibrium) that provide absolute MW information as well as column fractionation method using Mw/Mn measurements.
MW distribution analysis of SRM706
Analysis method |
Mw x 104 |
Mn x 104 |
Mw/Mn |
KF-406LHQ x 2
(4.6mmID*250mm) |
26.22 |
12.26 |
2.13 |
26.46 |
12.43 |
2.13 |
KF-806M
(8.0mmID*300mm) |
26.00 |
12.40 |
2.10 |
28.00 |
12.90 |
2.17 |
Light scattering |
25.78 |
- |
- |
Sedimentation equilibrum |
28.81 |
- |
- |
Column fractionation |
|
|
2.10 |