Features of RSpak DE Series (5) (Selection of HPLC System)
To realize the best performance of RSpak DE-413 column, we recommend to use the following HPLC system with the column.
Device | Recommended specifications and/or setting | Example |
---|---|---|
Pump | Flow rate: max. 1mL/min. | |
Injector | Low dead volume type Sample loop: max. 50micro-microL(5micro-L loop is recommended) |
|
From injector to column |
0.1 to 0.13mmID, 100mm length piping | |
From column to detector |
0.1mmID piping | |
Detector | Detector with a micro cell should be used. Response should be set to “FAST”. |
Shodex RI-104 (small cell volume) |
Sample : (A); 1.5micro-L, (B); 5micro-L
1. Methylethyl ketone 0.2%
2. Diethyl ketone 0.2%
3. Di-n-propyl ketone 0.4%
4. Di-n-butyl ketone 0.6%
Column : Shodex RSpak DE-413 (4.6mmID*150mm) Eluent : CH3CN/H2O=50/50 Flow rate : 1.0mL/min Detector : UV(262nm) Column temp. : 40deg-C