Calibration Curves for KW-800 Series
Calibration curves for PROTEIN KW-800 series (columns for aqueous SEC (GFC)) are shown.
Sample :
1. Thyroglobulin (MW: 669,000)
2. Aldolase (MW: 158,000)
3. BSA (MW: 67,000)
4. Ovalbumin (MW: 43,000)
5. Peroxidase (MW: 40,200)
6. Adenylate kinase (MW: 32,000)
7. Myoglobin (MW: 17,000)
8. Ribonuclease A (MW: 13,700)
9. Aprotinin (MW: 6,500)
10. Vitamin B12 (MW: 1,350)
Columns : Shodex PROTEIN KW-800 series (8.0mmID*300mm each) Eluent : 50mM Sodium phosphate buffer(pH7.0) + 0.3M NaCl Flow rate : 1.0mL/min Detector : UV(220nm) Column temp. : Room temp.