Column Test Conditions

To check the deterioration of column performance, measure the column plate number and compare it with the number written in the inspection data sheet which is attached to the column. The test conditions for calculation of the plate number is determined for each column type and shown in the table below.
When you first use a column, we recommend you to measure its plate number before use. Plate number (N) can be caluculated by the formula in the right figure.
For the columns whose test conditions are shown with blue letters, the column performance should be measured by resolution, and not plate number.

Column type
Test conditions
(1)Sample (2)Injection volume (3)Eluent (4)Flow rate (5)Temperature (6)Detector
ODP-50 6D
ODP-50 6E
ODP-50 4B
ODP-50 4D
ODP-50 4E
ODP-50 2D
(1)0.75micro-L/mL n-Hexyl benzoate(ODP-50 6D, ODP-50 6E, ODP-50 4B, ODP-50 4D, ODP-50 4E) 0.08% n-Hexyl Benzoate(ODP-50 2D) (2)10micro-L(ODP-50 6D), 17micro-L(ODP-50 6E), 3micro-L(ODP-50 4B), 6micro-L(ODP-50 4D, ODP-50 4E.), 7micro-L(ODP-50 2D), (3)CH3CN/H2O=65/35 (4)1.0mL/min(ODP-50 6D, ODP-50 6E), 0.6mL/min(ODP-50 4B, ODP-50 4D, ODP-50 4E), 0.2mL/min(ODP-50 2D) (5)30deg-C(ODP-50 6D, ODP-50 6E, ODP-50 4B, ODP-50 4D, ODP-50 4E), Ambient(ODP-50 2D) (6)UV(254nm)
ODP-40 4D
ODP-40 4E
(1)0.75micro-L/mL n-Hexyl benzoate (2)6micro-L (3)CH3CN/H2O=65/35 (4)0.6mL/min (5)30deg-C (6)UV(254nm)
C8P-50 4D
C8P-50 4E
(1)0.75micro-L/mL n-Hexyl benzoate (2)6micro-L(C8P-50 4D), 10micro-L(C8P-50 4E) (3)CH3CN/H2O=65/35 (4)0.6mL/min (5) 30deg-C (6)UV(254nm)
C4P-50 4D
C4P-50 4E
(1)0.75micro-L/mL n-Hexyl benzoate (2)6micro-L (3)CH3CN/H2O=65/35 (4)0.6mL/min (5) 30deg-C (6)UV(254nm)
ODP2 HP-4B
ODP2 HP-4D
ODP2 HP-4E
ODP2 HP-2B
ODP2 HP-2D
(1)1,000mg/L Toluene (2)5micro-L(ODP2 HP-4B, ODP2 HP-4D, ODP2 HP-4E), 2micro-L(ODP2 HP-2B), 1micro-L(ODP2 HP-2D) (3)H2O/CH3CN=55/45 (4)0.5mL/min(ODP2 HP-4B, ODP2 HP-4D, ODP2 HP-4E), 0.1mL/min(ODP2 HP-2B, ODP2 HP-2D) (5)40deg-C (6)UV(254nm)
RP18-415 (1)1% Acetone (2)3micro-L (3) CH3CN/H2O=95/5 (4)1.0mL/min (5)Ambient (6)UV(254nm)
RP18-613
RP18-413
(1)1% 3-Pentanone (2)5micro-L (3) CH3CN/THF/H2O=30/30/40 (4)0.8mL/min(RP18-613), 0.6mL/min(RP18-413) (5)Ambient (6)UV(262nm)
DS-613 (1)0.1% Benzene (2)20micro-L (3) CH3CN/THF/H2O=40/30/30 (4)0.8mL/min (5)Ambient (6)UV(254nm)
DS-413 (1)1% 3-Pentanone (2)5micro-L (3) CH3CN/THF/H2O=30/30/40 (4)0.6mL/min (5)Ambient (6)UV(262nm)
DE-613 (1)0.2% 1,5-Pentanediol (2)20micro-L (3) H2O (4)0.8mL/min (5)Ambient (6)RI
DE-413
DE-413L
DE-413S
(1)0.6% Di-n-butyl ketone (2)10micro-L(DE-413), 20micro-L(DE-413L), 8micro-L(DE-413S) (3) CH3CN/H2O=50/50 (4)1.0mL/min (5)40deg-C(DE-413, DE-413L), Ambient(DE-413S) (6)UV (262nm)
DE-213 (1)0.6% Di-n-butyl ketone (2)5micro-L (3) CH3CN/H2O=50/50 (4)0.2mL/min (5)40deg-C (6)UV (262nm)
GOLF-413
CARB-413
CARB-413L
(1)0.6% Di-n-butyl ketone (2)10micro-L (3) CH3CN/H2O=50/50 (4)1.0mL/min (5)40deg-C (6)UV (262nm)
DM-614 (1)2% Oxalic acid (2)5micro-L (3)5mM H3PO4 aq. (4)0.8mL/min (5)Ambient (6)RI
NN-814
NN-614
NN-414
(1)0.1% Uracil (2)10micro-L(NN-814), 7micro-L(NN-614, NN-414) (3)0.1M NaH2PO4(pH3.0) (4)1.0mL/min(NN-814), 0.8mL/min(NN-614), 0.3mL/min(NN-414) (5)35deg-C (6)UV(260nm)(NN-814, NN-614), RI(NN-414)
JJ-50 4D
JJ-50 2D
(1)1.0% Acetone (2)2micro-L(JJ-4D), 0.5micro-L(JJ-2D) (3)CH3CN/H2O=60/40 (4)1.0mL/min(JJ-4D), 0.15mL/min(JJ-2D) (5)Ambient (6)UV(280nm)
NH2P-50 4D
NH2P-50 4E
NH2P-50 2D
(1)5mg/mL Sucrose (2)6micro-L(NH2P-50 4D, NH2P-50 4E), 1.5micro-L(NH2P-50 2D) (3)CH3CN/H2O=75/25 (4)0.6mL/min(NH2P-50 4D), 1.0mL/min(NH2P-50 4E), 0.2mL/min(NH2P-50 2D) (5)30deg-C(NH2P-50 4D, NH2P-50 4E), 35deg-C(NH2P-50 2D) (6)RI
F-411/S
F-411
F-511
(1)0.15% t-Butyl benzene (2)20micro-L(F-411/S), 15micro-L(F-411, F-511) (3)CH3OH/H2O=80/20 (4)1.0mL/min (5)Ambient (6)UV(254nm)
C18M 4D
C18M 4E
C18P 4D
C18P 4E
(1)0.61mg/mL Naphthalene (2)1.5micro-L(C18M 4D, C18P 4D), 2micro-L(C18M 4E, C18P 4E) (3)CH3OH/H2O=70/30 (4)1.0mL/min (5)30deg-C (6)UV(254nm)
C18M 2D
C18M 2E
(1)3.05mg/mL Naphthalene (2)0.2micro-L (3)CH3CN/H2O=60/40 (4)0.2mL/min (5)30deg-C (6)UV(254nm)
E-411
5SIL 4D
5SIL 4E
(1)0.5mg/mL p-Nitrobenzyl alcohol (2)1.5micro-L(5SIL 4D), 2micro-L(5SIL 4E) (3)Hexane/C2H5OH=95/5 (4)1.0mL/min (5)30deg-C (6)UV(254nm)
5C8 4D
5C8 4E
(1)0.61mg/mL Naphthalene (2)1.5micro-L(5C8 4D), 2micro-L(5C8 4E) (3)CH3OH/H2O=66/34 (4)1.0mL/min (5)30deg-C (6)UV(254nm)
5C4 4D (1)0.61mg/mL Naphthalene (2)1.5micro-L (3)CH3OH/H2O=55/45 (4)1.0mL/min (5)30deg-C (6)UV(254nm)
5CN 4D
5CN 4E
(1)0.60mg/mL Naphthalene (2)1.5micro-L(5CN 4D), 2micro-L(5CN 4E) (3)CH3OH/H2O=40/60 (4)1.0mL/min (5)30deg-C (6)UV(254nm)
5NPE 4D (1)0.61mg/mL Naphthalene (2)1.5micro-L (3)CH3OH/H2O=55/45 (4)1.0mL/min (5)30deg-C (6)UV(254nm)
5NH 4D (1)3.47mg/mL p-Nitrophenyl-α-D-glucopyranoside (2)1.5micro-L (3)CH3CN/H2O=95/5 (4)1.0mL/min (5)30deg-C (6)UV(254nm)
To check the deterioration of column performance, measure the column plate number and compare it with the number written in the inspection data sheet which is attached to the column. The test conditions for calculation of the plate number is determined for each column type and shown in the table below.
When you first use a column, we recommend you to measure its plate number before use. Plate number (N) can be caluculated by the formula in the right figure.
For the columns whose test conditions are shown with blue letters, the column performance should be measured by resolution, and not plate number.

Column type
Test conditions
(1)Sample (2)Injection volume (3)Eluent (4)Flow rate (5)Temperature (6)Detector
SP-825 (1)0.5% Acetone (2)20micro-L (3)50mM Na2SO4 aq. (4)1.0mL/min (5)Ambient (6)UV(280nm)
SP-420N (1)0.2% Glycyl-L-tyrosine (2)1micro-L (3)20mM Acetate buffer + 0.5M Na2SO4(pH5.0) (4)1.0mL/min (5)Ambient (6)UV(280nm)
CM-825 (1)0.5% Acetone (2)20micro-L (3)50mM Na2SO4 aq. (4)1.0mL/min (5)Ambient (6)UV(280nm)
ES-502C 7C (1)0.025% Cytosine (2)15micro-L (3)100mM Sodium phosphate buffer(pH4.4) (4)1.0mL/min (5)30deg-C (6)UV(250nm)
P-421S (1)1% Ethylene glycol (2)4micro-L (3)H2O (4)0.5mL/min (5)Ambient (6)RI
QA-825 (1)0.5% Acetone (2)20micro-L (3)50mM Na2SO4 aq. (4)1.0mL/min (5)Ambient (6)UV(280nm)
DEAE-825 (1)0.5% Acetone (2)20micro-L (3)50mM Na2SO4 aq. (4)1.0mL/min (5)Ambient (6)UV(280nm)
DEAE3N-4T (1)200mg/L NaNO3 (2)5micro-L (3)30mM phosphate buffer(pH6.0) + 0.2M NaCl (4)1.0mL/min (5)Ambient (6)UV(210nm)
ES-502N 7C (1)0.025% Xanthine (2)10micro-L (3)50mM 1,3-Diaminopropane + 50mM NaCl(pH10.0) (4)1.0mL/min (5)30deg-C (6)UV(250nm)
WA-624 (1)0.03% UMP (2)10micro-L (3)0.1M Sodium phosphate buffer(pH3.0)/CH3CN=80/20 (4)1.0mL/min (5)Ambient (6)UV(260nm)
To check the deterioration of column performance, measure the column plate number and compare it with the number written in the inspection data sheet which is attached to the column. The test conditions for calculation of the plate number is determined for each column type and shown in the table below.
When you first use a column, we recommend you to measure its plate number before use. Plate number (N) can be caluculated by the formula in the right figure.
For the columns whose test conditions are shown with blue letters, the column performance should be measured by resolution, and not plate number.

Column type
Test conditions
(1)Sample (2)Injection volume (3)Eluent (4)Flow rate (5)Temperature (6)Detector
PH-814 (1)0.5% Acetone (2)20micro-L (3)H2O (4)1.0mL/min (5)Ambient (6)UV(280nm)

To check the deterioration of column performance, measure the column plate number and compare it with the number written in the inspection data sheet which is attached to the column. The test conditions for calculation of the plate number is determined for each column type and shown in the table below.
When you first use a column, we recommend you to measure its plate number before use. Plate number (N) can be caluculated by the formula in the right figure.
For the columns whose test conditions are shown with blue letters, the column performance should be measured by resolution, and not plate number.

Column type
Test conditions
(1)Sample (2)Injection volume (3)Eluent (4)Flow rate (5)Temperature (6)Detector
YS-50 (1)20mg/L Ca2+ (2)10micro-L (3)4mM Methanesulfonic acid aq. (4)1.0mL/min (5)40deg-C (6)CD
YK-421 (1)20mg/L Ca2+ (2)15micro-L (3)5mM Tartaric acid + 1mM Dipicolinic acid + 24mM Boric acid aq. (4)1.0mL/min (5)40deg-C (6)CD
Y-521 (1)20mg/L Na+ (2)50micro-L (3)4mM HNO3 aq. (4)1.5mL/min (5)40deg-C (6)CD
T-521 (1)20mg/L Na+ (2)20micro-L (3)3mM HNO3 aq. (4)1.0mL/min (5)40deg-C (6)CD
R-621 (1)1% Ethylene glycol (2)2micro-L (3)H2O (4)1.0mL/min (5)Ambient (6)RI
NI-424 (1)10mg/L SO42- (2)50micro-L (3)8mM 4-Hydroxybenzoic acid + 2.8mM Bis(2-hydroxyethyl)imino- tris(hydroxymethyl)methane + 2mM Phenylboronic acid + 5micro-M trans-1,2-Aminocyclohexane-N,N,N’,N’-4 acetic acid aq.(4)1.0mL/min (5)40deg-C (6)CD
I-524A (1)10mg/L NO3 (2)30micro-L (3)2.5mM Phthalic acid aq.(pH4.0) (4)1.5mL/min (5)40deg-C (6)CD
SI-90 4E (1)15mg/L SO42- (2)20micro-L (3)1.8mM Na2CO3 + 1.7mM NaHCO3 aq. (4)1.0mL/min (5)Ambient (6)Suppressed CD
SI-91 4C (1)10mg/L Br (2)20micro-L (3)1.8mM Na2CO3 + 1.7mM NaHCO3 aq. (4)1.0mL/min (5)40deg-C (6)Suppressed CD
SI-50 4E (1)10mg/L Br (2)20micro-L (3)3.2mM Na2CO3 + 1.0mM NaHCO3 aq. (4)0.7mL/min (5)Ambient (6)Suppressed CD
SI-52 4E (1)40mg/L SO4 (2)50micro-L (3)3.6mM Na2CO3 aq. (4)0.8mL/min (5)45deg-C (6)Suppressed CD

To check the deterioration of column performance, measure the column plate number and compare it with the number written in the inspection data sheet which is attached to the column. The test conditions for calculation of the plate number is determined for each column type and shown in the table below.
When you first use a column, we recommend you to measure its plate number before use. Plate number (N) can be caluculated by the formula in the right figure.
For the columns whose test conditions are shown with blue letters, the column performance should be measured by resolution, and not plate number.

Column type
Test conditions
(1)Sample (2)Injection volume (3)Eluent (4)Flow rate (5)Temperature (6)Detector
CDA-453 HQ (1)0.01% 6-Fluoro-tryptophan (2)10micro-L (3)H2O (4)0.3mL/min (5)20deg-C (6)UV(254nm)
CDB-453 HQ (1)0.01% DL-Met β-naphthylamide (2)10micro-L (3)1.7% TEA/Acetic acid(pH4.0) + 0.1M NaCl in 20% CH3CN + 80% H2O (4)0.2mL/min (5)20deg-C (6)UV(254nm)
CDC-453 HQ (1)0.02% Dancyl-leucine (2)20micro-L (3)1% Acetic acid + 0.2M NaCl in 10% CH3CN + 90% H2O (4)0.3mL/min (5)Ambient (6)UV(254nm)
CDBS-453 (1)0.1% Nefopam-HCl (2)10micro-L (3)CH3CN/(1% Acetic acid + 0.2M NaCl) aq.=30/70 (4)0.5mL/min (5)Ambient (6)UV(254nm)
CRX-853 (1)0.1% RS-Mandelic acid (2)20micro-L (3)0.25mM CuSO4 aq. (4)1.0mL/min (5)Ambient (6)UV(280nm)
ABA-894 (1)0.1% DL-Tryptophan (2)3micro-L (3)50mM Tris-HCl buffer + 0.15M(pH7.4) (4)1.0mL/min (5)Ambient (6)UV(280nm)

To check the deterioration of column performance, measure the column plate number and compare it with the number written in the inspection data sheet which is attached to the column. The test conditions for calculation of the plate number is determined for each column type and shown in the table below.
When you first use a column, we recommend you to measure its plate number before use. Plate number (N) can be caluculated by the formula in the right figure.
For the columns whose test conditions are shown with blue letters, the column performance should be measured by resolution, and not plate number.

Column type
Test conditions
(1)Sample (2)Injection volume (3)Eluent (4)Flow rate (5)Temperature (6)Detector
SC1011
SC1811
(1)1% Glycerol (2)5micro-L (3)H2O (4)1.0mL/min (5)80deg-C (6)RI
SP0810 (1)1% Glycerol (2)5micro-L (3)H2O (4)1.0mL/min (5)80deg-C (6)RI
KS-801
~KS-806
(1)1% Ethylene glycol(KS-801), 2.5% Ethylene glycol(KS-802), 0.5% Ethylene glycol(KS-803), 0.8% Glucose(KS-804, KS-806), 0.4% Glucose(KS-805) (2)5micro-L(KS-801, KS-802, KS-804, KS-805, KS-806), 15micro-L(KS-803) (3)H2O (4)1.0mL/min (5)50deg-C (6)RI
KS-807 (1)0.5% Glucose (2)5micro-L (3)H2O (4)1.0mL/min (5)25deg-C (6)RI
SC1211 (1)1% Glycerol (2)10micro-L (3)CH3CN/H2O=25/75 (4)1.0mL/min (5)50deg-C (6)RI
DC-613 (1)2% Sucrose (2)8micro-L (3)CH3CN/H2O=70/30 (4)0.8mL/min (5)50deg-C (6)RI
SZ5532 (1)1% Glucose (2)5micro-L (3)CH3CN/H2O=70/30 (4)1.0mL/min (5)50deg-C (6)RI
MN-431 (1)0.5% Mannitol (2)10micro-L (3)H2O (4)0.5mL/min (5)60deg-C (6)RI

To check the deterioration of column performance, measure the column plate number and compare it with the number written in the inspection data sheet which is attached to the column. The test conditions for calculation of the plate number is determined for each column type and shown in the table below.
When you first use a column, we recommend you to measure its plate number before use. Plate number (N) can be caluculated by the formula in the right figure.
For the columns whose test conditions are shown with blue letters, the column performance should be measured by resolution, and not plate number.

Column type
Test conditions
(1)Sample (2)Injection volume (3)Eluent (4)Flow rate (5)Temperature (6)Detector
SH1011
SH1821
(1)1% Glycerol (2)5micro-L (3)5mM H2SO4 aq. (4)1.0mL/min (5)50deg-C (6)RI
KC-811
KC-811 6E
(1)3% Acetic acid (2)5micro-L (3)0.1% H3PO4 aq. (4)1.0mL/min (5)40deg-C (6)RI

To check the deterioration of column performance, measure the column plate number and compare it with the number written in the inspection data sheet which is attached to the column. The test conditions for calculation of the plate number is determined for each column type and shown in the table below.
When you first use a column, we recommend you to measure its plate number before use. Plate number (N) can be caluculated by the formula in the right figure.
For the columns whose test conditions are shown with blue letters, the column performance should be measured by resolution, and not plate number.

Column type
Test conditions
(1)Sample (2)Injection volume (3)Eluent (4)Flow rate (5)Temperature (6)Detector
KW-802.5
to KW-804
(1)0.2% Ethylene glycol (2)30micro-L (3)H2O (4)1.0mL/min (5)Ambient (6)RI
KW402.5
to KW405
(1)0.05% Uridine(KW402.5-4F, KW403-4F) 0.02% Uridine(2)1micro-L (3)50mM Phosphate buffer + 0.3M NaCl(pH7.0)
(4)0.35mL/min (5)Ambient (6)UV(280nm)
SB-802 HQ
to SB-806M HQ
(1)1% Ethylene glycol(802 HQ), 0.5% Ethylene glycol(802.5 HQ to 806M HQ) (2)10micro-L(802 HQ), 15micro-L(802.5 HQ to 806M HQ) (3)H2O (4)1.0mL/min (5)Ambient (6)RI
SB-807 HQ (1)0.5% Glucose (2)10micro-L (3)H2O (4)1.0mL/min (5)Ambient (6)RI
SB401
to SB404
(1)0.5% Ethylene glycol (2)5micro-L (3)H2O (4)0.3mL/min (5)Ambient (6)RI
GS-220 HQ
GS-320 HQ
GS-520 HQ
GS-620 HQ
(1)1% Ethylene glycol (2)30micro-L (3)H2O (4)0.6mL/min (5)30deg-C (6)RI
GS-320 4B
GS-320 4D
GS-320 2D
(1)0.5% Ethylene glycol (2)2micro-L(GS-320 4B, GS-320 4D), 1.5micro-L(GS-320 2D) (3)H2O (4)0.25mL/min(GS-320 4B), 0.2mL/min(GS-320 4E), 0.05mL/min(GS-320 2D) (5)35deg-C (6)RI
GS-320 7E (1)1% Ethylene glycol (2)25micro-L (3)H2O (4)1.0mL/min (5)30deg-C (6)RI
GS-620 7G-P (1)1% Ethylene glycol (2)50micro-L (3)H2O (4)1.0mL/min (5)30deg-C (6)RI
GF-310 HQ
GF-510 HQ
GF-710 HQ
GF-7M HQ
(1)1% Ethylene glycol (2)30micro-L (3)H2O (4)0.6mL/min (5)30deg-C (6)RI
GF-310 4B
GF-310 4D
GF-310 4E
GF-310 2D
(1)0.5% Ethylene glycol (2)2micro-L(GF-310 4B, GF-310 4D), 3micro-L(GF-310 4E), 1micro-L(GF-310 2D) (3)H2O (4)0.25mL/min(GF-310 4B), 0.2mL/min(GF-310 4D, GF-310 4E), 0.05mL/min(GF-310 2D) (5)35deg-C (6)RI
KF-801
to KF-807L
(1)0.1% Propylbenzene (2)20micro-L (3)THF (4)1.0mL/min (5)Ambient (6)UV(254nm)
K-801
to K-807L
(1)0.1% Propylbenzene (2)20micro-L (3)Chloroform (4)1.0mL/min (5)Ambient (6)UV(254nm)
KD-801
to KD-807
(1)1.0% Acetone (2)20micro-L (3)DMF (4)1.0mL/min (5)Ambient (6)RI
KF-601
to KF-607
(1)0.5% Propylbenzene(KF-601), 0.1% Dicyclohexyl phthalate(KF-602 to KF-607) (2)5micro-L (3)THF (4)0.5mL/min (5)Ambient (6)UV(254nm)
KF-401HQ
to KF-406LHQ
(1)0.5% Propylbenzene(KF-401HQ), 0.1% Dicyclohexyl phthalate(KF-402HQ to KF-406LHQ) (2)5micro-L (3)THF (4)0.3mL/min (5)Ambient (6)UV(254nm)
HFIP-803
to HFIP-807
(1)5% Acetone (2)10micro-L (3)HFIP (4)1.0mL/min (5)Ambient (6)UV(280nm)
HFIP-603
to HFIP-607
(1)5% Acetone (2)2micro-L (3)HFIP (4)0.25mL/min (5)Ambient (6)UV(280nm)
HT-803
to HT-807
(1)10% Monochlorobenzene (2)20micro-L (3)Toluene (4)1.0mL/min (5)Ambient (6)RI
UT-802.5
UT-806M
UT-807
(1)10% Monochlorobenzene (2)10micro-L (3)Toluene (4)1.0mL/min (5)Ambient (6)RI
AT-806MS (1)1% Acetone (2)20micro-L (3)Toluene (4)1.0mL/min (5)Ambient (6)RI
LF-804
LF-604
LF-404
(1)0.1% Propylbenzene(LF-804), 0.1% Dicyclohexyl phthalate(LF-604, LF-404) (2)20micro-L(LF-804), 5micro-L(LF-604, LF-404) (3)THF (4)1.0mL/min(LF-804), 0.5mL/min(LF-604), 0.3mL/min(LF-404) (5)Ambient (6)UV(254nm)