Monovalent and Divalent Cation Standards (6) (Comparison of YS-50 and YK-421)
The six main cations can be separated using a newly developed column, IC YS-50. Comparing with the conventional column IC YK-421, the theoretical plate number of divalent cations becomes more than double and the separation of Na+ and NH4+ becomes about 1.2 times.
Sample :
1. Li+ 2ppm
2. Na+ 10ppm
3. NH4+ 10ppm
4. K+ 20ppm
5. Mg2+ 10ppm
6. Ca2+ 20ppm
Columns : Shodex IC YS-50 (4.6mmID*125mm) Shodex IC YK-421 (4.6mmID*125mm) Eluent : 4mM Methanesulfonic acid aq. 5mM Tartaric acid + 1mM Dipicolinic acid + 1.5g/L Boric acid aq. Flow rate : 1.0mL/min 1.0mL/min Detector : Non-suppressed conductivity Non-suppressed conductivity Column temp. : 40deg-C 40deg-C