Ion Chromatography Columns
To check the deterioration of column performance, measure the column plate number and compare it with the number written in the inspection data sheet which is attached to the column. The test conditions for calculation of the plate number is determined for each column type and shown in the table below.
When you first use a column, we recommend you to measure its plate number before use. Plate number (N) can be caluculated by the formula in the right figure.
For the columns whose test conditions are shown with blue letters, the column performance should be measured by resolution, and not plate number.
Column type |
Test conditions
(1)Sample (2)Injection volume (3)Eluent (4)Flow rate (5)Temperature (6)Detector |
---|---|
YS-50 | (1)20mg/L Ca2+ (2)10micro-L (3)4mM Methanesulfonic acid aq. (4)1.0mL/min (5)40deg-C (6)CD |
YK-421 | (1)20mg/L Ca2+ (2)15micro-L (3)5mM Tartaric acid + 1mM Dipicolinic acid + 24mM Boric acid aq. (4)1.0mL/min (5)40deg-C (6)CD |
Y-521 | (1)20mg/L Na+ (2)50micro-L (3)4mM HNO3 aq. (4)1.5mL/min (5)40deg-C (6)CD |
T-521 | (1)20mg/L Na+ (2)20micro-L (3)3mM HNO3 aq. (4)1.0mL/min (5)40deg-C (6)CD |
R-621 | (1)1% Ethylene glycol (2)2micro-L (3)H2O (4)1.0mL/min (5)Ambient (6)RI |
NI-424 | (1)10mg/L SO42- (2)50micro-L (3)8mM 4-Hydroxybenzoic acid + 2.8mM Bis(2-hydroxyethyl)imino- tris(hydroxymethyl)methane + 2mM Phenylboronic acid + 5micro-M trans-1,2-Aminocyclohexane-N,N,N’,N’-4 acetic acid aq.(4)1.0mL/min (5)40deg-C (6)CD |
I-524A | (1)10mg/L NO3– (2)30micro-L (3)2.5mM Phthalic acid aq.(pH4.0) (4)1.5mL/min (5)40deg-C (6)CD |
SI-90 4E | (1)15mg/L SO42- (2)20micro-L (3)1.8mM Na2CO3 + 1.7mM NaHCO3 aq. (4)1.0mL/min (5)Ambient (6)Suppressed CD |
SI-91 4C | (1)10mg/L Br– (2)20micro-L (3)1.8mM Na2CO3 + 1.7mM NaHCO3 aq. (4)1.0mL/min (5)40deg-C (6)Suppressed CD |
SI-50 4E | (1)10mg/L Br– (2)20micro-L (3)3.2mM Na2CO3 + 1.0mM NaHCO3 aq. (4)0.7mL/min (5)Ambient (6)Suppressed CD |
SI-52 4E | (1)40mg/L SO4– (2)50micro-L (3)3.6mM Na2CO3 aq. (4)0.8mL/min (5)45deg-C (6)Suppressed CD |