Eluent Conditions for KW400-4F and PROTEIN KW-800

Eluent conditions for KW400-4F series and KW-800 series columns

1) Typical eluent
The following solutions are generally used as the eluent.
Phosphate buffer solution
Tris hydrochloric solution
Acetate buffer solution

2) Salt to be added
Salts such as Na2SO4, K2SO4 and (NH4)2SO4 are usually added to the above mentioned buffer solutions. The suitable salt concentration is 0.1 to 0.3M.
CAUTlON
(1) The pH range of the eluent should be 3.0 to 7.5 range.
(2) Use of salts containing chlorine ions should preferably be avoided, as chlorine ions are corrosive to the column or piping of the instrument.

3) Use of urea or guanidlne hydrochloride
Urea or 6M guanidine hydrochloride solutions, which are often used as modifier of proteins, can be used as the eluent.
CAUTION
Since the replacement of the eluent takes a fairly long time, it is recommended to use one column specifically for such an eluent.

4) Use of surfactant
Aqueous solutions of surfactant such as SDS and Brij can also be used.

5) Use of polar organic solvent
Polar organic solvents, such as acetonitrile, methanol and ethanol can also be used as the eluent.