Standard Proteins (21) (SP-420N)
Six kinds of protein standards were analyzed using IEC SP-420N (a column for strong cation exchange chromatography packed with non-porous gel).
Sample :
1. Myoglobin
2. Trypsinogen
3. Ribonuclease A
4. alpha Chymotrypsinogen A
5. Cytochrome c
6. Lysozyme
Column : Shodex IEC SP-420N (4.6mmID*35mm) Eluent : (A); 20mM Sodium phosphate buffer(pH7.0) (B); (A) + 0.5M NaCl Linear gradient: 0min to 10min, (A) to (B) Flow rate : 1.5mL/min Detector : UV(280nm) Column temp. : Room temp.