High Sensitive Analysis of Metal Bonded Amino Acids
High sensitive analysis of metal bonded amino acids was performed using GS320A-M8E (0.8mmID x 250mm) and ICP-MS. GS320A-M8E is a micro-column of Asahipak GS-320 HQ. Using the micro column, It is possible to reduce the injection volume and eluent to 1/200 and 1/100 compared with the case where a conventional column is used, respectively.
Sample :
MeSeCys (Methylselenocysteine)
SeMet (Selenomethionine)
(Left) (Right) LC : capillary HPLC normal HPLC (Tubing: 0.03mmID fused-silica capillary tube) (Tubing: 0.25mmID PEEK tube) Columns : Shodex GS320A-M8E (0.8mmID*250mm) Shodex Asahipak GS-320 HQ (7.5mmID*300mm) Eluent : 50mM Ammonium acetate(pH6.5) 50mM Ammonium acetate(pH6.5) Flow rate : 5.5micro-L/min 500micro-L/min Detector : ICP-MS Se at m/z=82 ICP-MS Se at m/z=82
Courtesy of Dr. Yasumitsu Ogra, Graduate School of Pharmaceutical Sciences, Chiba University