Cyanide
Combining IEC DEAE-825 (a column for weak anion exchange chromatography) with an electrochemical detector (ECD) has permitted sensitive detection of cyanide. With this method, the detection limit of cyanide is 0.2 micro-g/L.
Sample : CN–
Column : Shodex IEC DEAE-825 (8.0mmID*75mm) Eluent : (10mM Sodium carbonate + 1mM Ethylenediamine) aq./CH3OH=90/10 Flow rate : 1.0mL/min Detector : ECD(Electrode; Ag 0mV SCE) Column temp. : 40deg-C
Total cyanide in river water was analysed using DEAE-825 (a column for weak anion exchange chromatography) and an electrochemical detector (ECD).
Sample : 100micro-L of distilled river water (water)
Column : Shodex IEC DEAE-825 (8.0mmID*75mm) Eluent : (10mM Sodium carbonate + 1mM Ethylenediamine)/CH3OH=90/10 Flow rate : 1.0mL/min Detector : ECD (Electrode; Ag, 0mV SCE) Column temp. : 40deg-C
RSpak KC-811 6E is suitable for the analysis of cyanide ion and cyanogen chloride, fulfills required measurement for drinking-water regulation in Japan.
Sample : 10micro-g/L each, 100micro-L
CN–, Cyanide ion
CNCl, Cyanogen chloride
Column : Shodex RSpak KC-811 6E (6.0mmID*250mm) Eluent : 1.0mM H2SO4 aq. Reagent A : Chloramine T solution Reagent B : 4-Pyridinecarboxylic acid-Pyrazolone solution Flow rate : (Eluent); 1.0mL/min (Reagent); 0.5mL/min each Detector : VIS(638nm) Column temp. : 40deg-C Reaction temp. : (Reagent A); 40deg-C (Reagent B); 80deg-C