GPC Clean-up of Pesticides Residuals in Foods

EV · Suitable for fractionation of residual pesticides in food
· Used in the purification preocess of test solution preparation method in Shoku-Au No.1003001 (October 3ed, 2006. Japan) of thw Pharmaceutical and Food Safety Bureau, MHLW, Section 2 “Simultaneous GC/MS (LC/MS) Analysis of Agricultual Chemicals in Livestock and Marine Products”
PAE · Suitable for cleaning up water samlpes such as blood and bottom sediment
· Highly effective for Fractionation of endocrine disruptors in environmental samples

(GPC clean-up for residual pesticides in food, etc)

Product code
Product name
Particle size
(micron)
Size(mm)
ID x Length
F6090005 CLNpak EV-200 16 2.0 x 150
F6090001
F6090002
CLNpak EV-2000
CLNpak EV-G
16
16
20.0 x 300
20.0 x 100
F6090003
F6090004
CLNpak EV-2000 AC
CLNpak EV-G AC
16
16
20.0 x 300
20.0 x 100
Product
name
Max. sample
load (mL)
Max. usable
pressure
(MPa)
Recommended
flow rate
(mL/min)
Shipping
solvent
EV-200
0.2
2.5 0.05
Ethylacetate/Cyclohexane=3/7
EV-2000
EV-G
5
2.5
4.0 to 5.0
Ethylacetate/Cyclohexane=3/7
Ethylacetate/Cyclohexane=3/7
EV-2000 AC
EV-G AC
5
2.5
4.0 to 5.0
Acetone/Cyclohexane=3/7
Acetone/Cyclohexane=3/7

Usable eluent : Acetone/Cyclohexan=1/9 to 5/5, Ethylacetate/Cyclohexane=1/9 to 5/5

(GPC clean-up for phthalic acid esters in sediments, bioogical samples, blood, etc)

Product code
Product name
Particle size
(micron)
Size(mm)
ID x Length
F7600025
F6810022
F6714007
CLNpak PAE-800
CLNpak PAE-2000
CLNpak PAE-G
5
5
9
8.0 x 300
20.0 x 300
8.0 x 50
F7600026
F6810023
F6714008
CLNpak PAE-800 AC
CLNpak PAE-2000 AC
CLNpak PAE-G AC
5
5
9
8.0 x 300
20.0 x 300
8.0 x 50
Product
name
Max. sample
load(mL)
Max. usable
pressure
(MPa)
Recommended
flow rate
(mL/min)
Shipping
solvent
PAE-800
PAE-2000
PAE-G
0.5
3
4.0
5.0
0.8
3.5
Acetonitrile
Acetonitrile
Acetonitrile
PAE-800 AC
PAE-2000 AC
PAE-G AC
0.5 (water:10%)
3 (water:10%)
4.0
5.0
0.8
3.5
Acetone
Acetone
Acetone

Usable eluent : Acetone, Acetonitrile

Recently, the analysis of pesticides residues in many types of foods have become more important because of the increasing demand from EPA regulatory agencies.
The GPC (Gel Permeation Chromatography) clean-up method is widely used to remove lipids, polymers, pigments and other high molecular weight compounds from a wide range of sample matrices including food, tissues, grains, plants and environmental samples. Using this method combined with GC/MS, GC or HPLC, it is possible to detect many pesticides residues with good reproducibility in a relatively short period of time.
Dichloromethane has been used as an eluent for GPC clean-up, but as it may itself cause environmental pollution. To overcome this, the possibility of using a non-halogenated solvent such as ethyl acetate/cyclohexane as eluent was tested recently, but compared with dichloromethane the separation efficiency was inferior.
CLNpak EV-2000 series for GPC clean-up is developed to be used with non-halogenated solvents such as ethyl acetate/cyclohexane as the eluent. Using such solvents, EV-2000 series can obtain almost the same results when dichloromethane is used. This is the unique feature of EV-2000 series.

Analytical condition is following. You do not need special system. You can use CLNpak EV-2000 in usual HPLC system. Please use eluent of HPLC grade and analysis grade for pesticide residues.

Elution profile for pesticides when CLNpak EV-2000 AC is used is shwn.

Elution volume of pesticides
(total 100%)
60
to
70
mL
70
to
80
mL
80
to
90
mL
90
to
100
mL
100
to
110
mL
110
to
120
mL
120
to
130
mL
130
to
140
mL
fluvalinate 16 84
chinomethionate 100
Pyrethroid cyhalothrin 100
flucythrinate 64 36
cypermethrin 28 72
cyfluthrin 18 82
deltamethrin 100
fenvalerate 100
Organo-
phosphorus
terbufos 4 96
diazinon 100
prothiofos 98 2
pirimiphos methyl 96 4
chlorpyrifos 80 20
isofenphos 33 67
malathion 29 71
parathion 100
phosalone 98 2
phenthoate 95 5
thiometon 61 39
EPN 7 93
fenitrothion 98 2
fenthion 95 5
dimethoate 94 6
acephate 75 25
methamidophos 100
N-contg. pretilachlor 2 16 82
pyridaben 25 75
chlorpropham 100
propanil 100
bitertanol 98 2
fenobucarb 96 4
mepronil 90 10
isoprocarb 72 28
aldicarb 100
ethiofencarb 100
iprodione 100
bendiocarb 96 4
imibenconazole 60 40
pirimicarb 53 47
carbaryl 61 35 4
mefenacet 97 3
Organo-
chlorine
dicofol 100
vinclozolin 100
aldrin 96 4
beta-HCH 86 14
alpha-HCH 80 20
pp’-DDT 72 28
pp’-DDE 49 51
procymidone 22 78
endosulfan 17 83
methoxychlor 3 42 55
dieldrin 95 5
endrin 89 11
captafol 100
captan 48 52
TPN 100
phthalide 90 10
Columns      : Shodex CLNpak EV-G AC (20mmID*100mm) + EV-2000 AC (20mmID*300mm)
Eluent       : Acetone/Cyclohexane=20/80
Flow rate    : 4.0mL/min
Column temp. : 40deg-C

Instead of dichloromethane which may cause the environmental pollution, non-halogenated solvent such as ethyl acetate and acetone can be used as eluent. Each solvent is mixed with cyclohexane and used for CLNpak EV-2000 series as the eluent. The effect of the solvents and their mixing ratio is shown below.

Sample :
1. 125mg/mL Corn oil
2. 1mg/2mL beta-Carotene
3. 1mg/2mL Fluvalinate
4. 0.3mg/2mL Chinomethionate

Columns      : (Left); Shodex CLNpak EV-G AC (20mmID*100mm) + EV-2000 AC (20mmID*300mm) 
               (Center); Shodex CLNpak EV-2000 (20mmID*300mm)
               (Right); Shodex CLNpak EV-G (20mmID*100mm) + EV-2000 (20mmID*300mm)
Eluent       : (Left); Acetone/Cyclohexane
               (Center); Ethyl acetate/Cyclohexane
               (Right); Dichloromethane/Cyclohexane
Flow rate    : 4.0mL/min
Detector     : UV(254nm)
Column temp. : 25deg-C

CLNpak EV-2000 AC is used for GPC clean-up of pesticides residuals in food with acetone/cyclohexane eluent.

Sample :
1. Palmitic acid, 2. Oleic acid
3. Linolenic acid, 4. Fluvalinate
5. Chinomethionate, 6. beta-Carotene
7. Methoprene, 8. Cyhalothrin
9. Cyfluthrin, 10. Diazinon
11. Malathion, 12. Chloropyrifos-methyl
13. Corn oil, 14. Bis(2-ethylhexyl) phthalate
15. Methoxychlor, 16. Perylene
17. Sulfur

Columns      : Shodex CLNpak EV-G AC (20mmID*100mm) + EV-2000 AC (20mmID*300mm) 
Eluent       : Acetone/Cyclohexane=20/80
Flow rate    : 5.0mL/min
Detector     : UV(254nm)
Column temp. : 25deg-C

CLNpak EV-2000 is used for GPC clean-up of pesticides residuals in food with ethyl acetate/cyclohexane eluent.

Sample :
1. Palmitic acid, 2. Oleic acid
3. Linolenic acid, 4. Fluvalinate
5. Chinomethionate, 6. beta-Carotene
7. Methoprene, 8. Cyhalothrin
9. Cyfluthrin, 10. Diazinon
11. Malathion, 12. Chloropyrifos-methyl
13. Corn oil, 14. Bis(2-ethylhexyl) phthalate
15. Methoxychlor, 16. Perylene
17. Sulfur

Columns      : Shodex CLNpak EV-G (20mmID*100mm) + EV-2000 (20mmID*300mm) 
Eluent       : Ethyl acetate/Cyclohexane=20/80
Flow rate    : 5.0mL/min
Detector     : UV(254nm)
Column temp. : 25deg-C

Removing pigment of spinach with CLNpak EV-2000 AC is shown.

Sample : Spinach

Columns      : CLNpak EV-G AC (20mmID*100mm) + EV-2000 AC (20mmID*300mm)
Eluent       : Acetone/Cyclohexane=20/80
Flow rate    : 5.0mL/min
Detector     : UV(254nm)
Column temp. : 25deg-C

Recovery of pesticides residuals in spinach when CLNpak EV-2000 AC is used for GPC Clean-up of pesticides residuals is shown.

Sample :
1. Corn oil
2. Fluvalinate
3. Chinomethionate

Injection
as spinach
Injection
dry weight
40 to 60mL 60 to 130mL Total
recovery
2.5g/5mL 9.6mg 9.3mg(97%) 0.9mg(9%) 106%
5g/5mL 19.3mg 16.6mg(86%) 2.0mg(10%) 96%
20g/5mL 77.0mg 68.5mg(89%) 10.0mg(13%) 102%
Columns      : Shodex CLNpak EV-G AC (20mmID*100mm) + EV-2000 AC (20mmID*300mm) 
Eluent       : Acetone/Cyclohexane=20/80
Flow rate    : 5.0mL/min
Detector     : UV(254nm)
Column temp. : 25deg-C

Effect of sample injection volume when CLNpak EV-2000 AC is used for GPC clean-up of pesticides residuals in food is shown.

Sample :
1. Corn oil
2. Fluvalinate

Column       : Shodex CLNpak EV-G AC (20mmID*100mm) + EV-2000 AC (20mmID*300mm)
Eluent       : Acetone/Cyclohexane=20/80
Flow rate    : 5.0mL/min
Detector     : UV(254nm)
Column temp. : 25deg-C

Effect of flow rate when CLNpak EV-2000 AC is used for GPC clean-up of pesticides residuals in food is shown.

Sample :
1. Corn oil
2. Bis(2-ethylhexyl) phthalate
3. Methoxychlor
4. Perylene
5. Sulfur

Columns      : Shodex CLNpak EV-G AC (20mmID*100mm) + EV-2000 AC (20mmID*300mm) 
Eluent       : Acetone/Cyclohexane=20/80
Detector     : UV(254nm)
Column temp. : 25deg-C

Effect of column temperature when CLNpak EV-2000 AC is used for GPC clean-up of pesticides residuals in foods is shown.

Sample :
1. Corn oil
2. Fluvalinate
3. Chinomethionate

Columns      : CLNpak EV-G AC (20mmID*100mm) + EV-2000 AC (20mmID*300mm)
Eluent       : Acetone/Cyclohexane=20/80
Flow rate    : 5.0mL/min
Detector     : UV(254nm)

The in-column solvent of CLNpak EV-2000 can be replaced easily without affecting column performance.

The column shipping solvent of EV-2000 column is ethyl acetate/cyclohexane(3/7).
(1) After pumping acetone/cyclohexane (2/8) at 2mL/min for 2h.
(2) After pumping ethyl acetate/cyclohexane (2/8) at 2mL/min for 2h.
(3) After pumping acetone/cyclohexane (2/8) at 2mL/min for 2h.
(4) After pumping ethyl acetate/cyclohexane (2/8) at 2mL/min for 2h.

Sample :
1. Corn oil
2. Fluvalinate
3. Chinomethionate

Columns      : Shodex CLNpak EV-G (20mmID*100mm) + EV-2000 (20mmID*300mm) 
Flow rate    : 4.0mL/min
Detector     : UV(254nm)
Column temp. : 25deg-C

The in-column solvent of CLNpak EV-2000 can be replaced easily without affecting column performance.
The column shipping solvent of EV-2000 column is ethyl acetate/cyclohexane(3/7).
(1) After pumping dichloroethane/cyclohexane(3/7) at 2mL/min for 2h.
(2) After pumping acetone/cyclohexane(2/8) at 2mL/min for 2h.
(3) After pumping dichloroethane/cyclohexane(3/7) at 2mL/min for 2h.
(4) After pumping acetone/cyclohexane(2/8) at 2mL/min for 2h.

Sample :
1. Corn oil
2. Fluvalinate
3. Chinomethionate

Columns      : Shodex CLNpak EV-G (20mmID*100mm) + EV-2000 (20mmID*300mm) 
Flow rate    : 4.0mL/min
Detector     : UV(254nm)
Column temp. : 25deg-C

The gel used for CLNpak EV-2000 AC has very good lot-to-lot reproducibility. The calibration curves of six different gel lots have almost same. The comparison of chromatograms of two lots among the six lots is also shown.

Sample :
1. beta-Carotene
2. Fluvalinate
3. Chinomethionate
4. Corn oil
5. Bis(2-ethylhexyl) phthalate
6. Methoxychlor
7. Perylene
8. Sulfur

Columns      : Shodex CLNpak EV-G AC (20mmID*100mm) + EV-2000 AC (20mmID*300mm)
Eluent       : Acetone/Cyclohexane=30/70
Flow rate    : 4.0mL/min
Detector     : UV(254nm)
Column temp. : 25deg-C

Swelling data of CLNpak EV gel is shown. Swell of EV gel is the almost same with many kind of solvents. It means EV gel can be used with many kind of solvents safely.

Solvent and swelled volume of EV-2000 gel

The data of nother GPC gel for comparison

Solvents Degree of swelling (%)
EV-2000
Other GPC gel
1. 100% Acetone
88
58
2. Acetone/Cyclohexane=50/50
94
82
3. Acetone/Cyclohexane=20/80
98
78
4. 100% Ethyl acetate
95
73
5. Ethyl acetate/Cyclohexane=50/50
97
85
6. Ethyl acetate/Cyclohexane=20/80
94
77
7. 100% Dichloromethane
100
100
8. Dichloromethane/Cyclohexane=50/50
104
103
9. Dichloromethane/Cyclohexane=30/70
102
93
10. Toluene/Cyclohexane=30/70
101
81

Ministry of Health, labour and welfare in Japan regulates maximum residue limits for agricultural chemicals in foods. Based on the revision of the “Food Sanitation Act” in 2003, “positive list system” was enacted May 29, 2006. In Food Sanitation Law #1003001 (Oct 3, 2006), section two states “Analytical method with GC/MS for multi-residues” and “Analytical method with LC/MS for multi-residues” in crops and animal products. In the method, gel permeation chromatography (GPC) has been adapted for the sample purification process. CLNpak EV-2000 AC is a column especially designed for the GPC clean up purpose. Below is the outline for the analytical method with GC/MS and LC/MS analysis for multi-residues.

Ministry of Health, labour and welfare in Japan regulates maximum residue limits for agricultural chemicals in foods. Based on the revision of the “Food Sanitation Law” in 2003, “positive list system” was enacted May 29, 2006. In Food Sanitation Law #1003001 (Oct 3, 2006), section two states “Analytical method with GC/MS for multi-residues” and “Analytical method with LC/MS for multi-residues” in crops and animal products. In the method, acrinathrin and tricyclazole were used as fractionation standards. For the purification of pesticide residuals in animal products, a GPC clean-up column, CLNpak EV-2000 AC is recommended.
Fractionation area is depending on the type of samples used for the GPC clean-up.

Sample : 5mL
Acrinathrin
Tricyclazole

Column       : Shodex CLNpak EV-G AC (20.0mmID*100mm) + EV-2000 AC (20.0mmID*300mm)
Eluent       : Acetone/Cyclohexane=1/4
Flow rate    : 5mL/min
Detector     : UV(254nm)
Column temp. : 40deg-C

Fluvalinate and chinomethionate were previously used as fractionation standards. In Food Sanitation Law #1003001 (Oct 3, 2006), section two states “Analytical method with GC/MS for multi-residues” and “Analytical method with LC/MS for multi-residues” in crops and animal products. In the method, acrinathrin and tricyclazole were used as fractionation standards. Compared to the fluvalinate and chinomethionate fractionation area, the one obtained by acrinathrin and tricyclazole is wider.

Sample : 40mg/L each, 5mL
Acrinathrin
Tricyclazole
Fluvalinate
Chinomethionate

Column       : Shodex CLNpak EV-G AC (20.0mmID*100mm) + EV-2000 AC (20.0mmID*300mm)
Eluent       : Acetone/Cyclohexane=1/4
Flow rate    : 5mL/min
Detector     : UV(254nm)
Column temp. : 40deg-C

Elution volume for lipid and fatty acids were monitored by CLNpak EV-2000 AC. Lipid and fatty acids elute fast, thus by fractionating between acrinathrin and tricyclazol, most lipid and fatty acids can be removed.

Sample : 5mL
Beef tallow
Acrinathrin
Tricyclazole
Oleic acid
Palmitic acid
Linolenic acid
Linoleic acid

Column       : Shodex CLNpak EV-G AC (20.0mmID*100mm) + EV-2000 AC (20.0mmID*300mm)
Eluent       : Acetone/Cyclohexane=1/4
Flow rate    : 5mL/min
Detector     : UV(254nm)
Column temp. : 40deg-C

Two different temperature settings (room temerature and 40 deg-C) were compared for the GPC clean-up method using CLNpakEV-2000 AC. At room temperature, the high viscosity causes peak broadening and increases the analysis time. Therefore, analysis at 40 deg-C is preferred. In Food Sanitation Law #1003001 (Released by Ministry of Health, labour and welfare in Japan, 2006), analysis temperature is set at 40 deg-C.

Sample : 5mL
Fatty tuna
Acrinathrin 40mg/L
Tricyclazole 40mg/L

Column       : Shodex CLNpak EV-G AC (20.0mmID*100mm) + EV-2000 AC (20.0mmID*300mm)
Eluent       : Acetone/Cyclohexane=1/4
Flow rate    : 5mL/min
Detector     : UV(254nm)

Main separation mode for the GPC clean-up is size exclusion mode: based on analytes’ size differences, analytes are eluted in the order of their sizes. However, by optimizing the eluent composition, partition/adsorption mode can be also used. Change in acetone and cyclohexane ratio does not influence elution volume of lipid much. However by decreasing acetone, retention time of pesticides increases. At acetone/cyclohexane=3/7, lipid and acrinathirin co-elutes. In Food Sanitation Law #1003001 (Released by Ministry of Health, labour and welfare in Japan, 2006), the ratio of acetone/cyclohexane is set at 1/4.

Sample : 40mg/L each, 5mL
Acrinathrin
Tricyclazole

Column       : Shodex CLNpak EV-G AC (20.0mmID*100mm) + EV-2000 AC (20.0mmID*300mm)
Eluent       : Acetone/Cyclohexane
Flow rate    : 5mL/min
Detector     : UV(254nm)
Column temp. : 40deg-C