Hydrophobic Interaction Analysis
| Recovery of proteins | |
|---|---|
| Protein | Recovery(%) |
| Cytochrome c | 104 |
| Myoglobin | 93 |
| Ribonuclease A | 100 |
| Ovalbumin | 103 |
| Lysozyme | 100 |
| alpha Chymotrypsin | 99 |
| alpha Chymotrypsinogen A | 96 |
| Recovery of enzyme activity |
|
|---|---|
| Enzyme | Recovery(%) |
| Trypsin | 96 |
| Hexokinase | 94 |
| Catalase | 88 |
| Lysozyme | 100 |
| Lipoxidase | 100 |
Sample load : 0.5mg each
Column : Shodex HIC PH-814
Effect of Sample load on peak width was measued using HIC PH-814 ( a hydrophobic interaction column ). A sample load of 0.4mg does not widen the peak. Even when injecting a higher load of 1mg sample, the peak only widens 1.3 times.

Sample :
1. alpha Chymotrypsinogen
2. Ribonuclease
3. Myoglobin
Column : Shodex HIC PH-814 (8.0mmID*75mm)
Eluent : (A); (B) + 1.8M Ammonium sulfate
(B); 0.1M Phosphate buffer(pH7.0)
Linear gradient: 0min to 60min, (A) to (B)
Flow rate : 1.0mL/min
Effect of salt concentration was measured using HIC PH-814 ( a hydrophobic interaction column ).

Sample :
1. alpha Chymotrypsinogen A
2. alpha Chymotrypsin
3. Lysozyme
4. Ovalbumin
5. Ribonuclease A
6. Myoglobin
7. Cytochrome c
Column : Shodex HIC PH-814 (8.0mmID*75mm)
Eluent : (A): (B) + Ammonium sulfate
(B): 0.1M Phosphate buffer(pH7.0)
Linear gradient: (A) to (B)
Flow rate : 1.0mL/min
Six kinds of protein standards were analyzed using HIC PH-814 ( a hydrophobic interaction column ).

Sample :
1. Cytochrome c
2. Myoglobin
3. Ribonuclease A
4. Ovalbumin
5. Lysozyme
6. alpha Chymotrypsinogen
Column : Shodex HIC PH-814 (8.0mmID*75mm)
Eluent : (A); (B) + 1.8M Ammonium sulfate
(B); 0.1M Phosphate buffer(pH7.0)
Linear gradient: 0min to 60min, (A) to (B)
Flow rate : 1.0mL/min
Detector : UV(280nm)
Column temp. : Room temp.
Lipoxidase from soybean was analyzed using HIC PH-814 ( a hydrophobic interaction column ). Three large peaks were obtained. Enzyme activity was found at the sosition of arrow mark after 40 minutes.

Sample : Lipoxidase from soybean
Column : Shodex HIC PH-814 (8.0mmID*75mm)
Eluent : (A); (B) + 1.8M Ammonium sulfate
(B); 0.1M Phosphate buffer(pH7.0)
Linear gradient: 0min to 60min, (A) to (B)
Flow rate : 1.0mL/min
Detector : UV(280nm)
Column temp. : 25deg-C
Two trypsins, alpha and beta were separated using HIC PH-814 ( a hydrophobic interaction column ).

Sample :
1. alpha Trypsin
2. Trypsinogen
3. beta Trypsin
Column : Shodex HIC PH-814 (8.0mmID*75mm)
Eluent : (A); (B) + 1.8M Ammonium sulfate
(B); 0.1M Phosphate buffer(pH3.0)
Linear gradient: 0min to 60min, (A) to (B)
Flow rate : 1.0mL/min
Detector : UV(280nm)
Column temp. : 25deg-C