Feature of High Temperature Analysis (1) (High Resolution)
ET-RP1 4D is polymer base column for high temperature analysis. Figure shows comparison of reduced plate height and analysis time between 40deg-C and 150deg-C. Smaller value of reduced plate height means better column resolution. The reduced plate height of acetophenone and benzene decreased at the 150deg-C. Mobile phase viscosity decreases at higher temperature and thus enhanced diffusion rate between the eluent and the polymeric stationary phase. Then resolution is improved.
Sample :
1. Uracil
2. Pyridine
3. Acetophenone
4. Benzene
(notes) Eluent was introduced to column after preheating and introduced to detector after cooling.
Column | : Shodex ET-RP1 4D (4.6mmID*150mm) | : Shodex ET-RP1 4D (4.6mmID*150mm) |
Eluent | : H2O/CH3CN=50/50 | : H2O/CH3CN=75/25 |
Flow rate | : 0.5mL/min | : 2.4mL/min |
Detector | : Diode array (210nm) | : Diode array (210nm) |
Column temp. | : 40deg-C | : 150deg-C |
*Column oven : Polaratherm 9000 Series (SandraSelerity Technologies, Inc)
Coutesy of Research Institude for Chromatography bvba