Recovery of Proteins and Enzime Activity (QA-825)
Recoveries of proteins and enzime activity for IEC QA-825 (a column forstrong anion exchange chromatography) are shown. When 0.5mg of each proteins was injected into the column by linear gradient elution, in which salt concentration increases linearly over one minute, sample recovery was found to be good. Then the same amount of enzyme was injected into the column in the same manner as the above protein and its recovery was also found to be good.
Recovery of proteins | |
---|---|
Protein | Recovery (%) |
Ovalbumin | 106 |
Conalbumin | 100 |
beta- Lactoglobulin | 99 |
Trypsin inhibitor | 99 |
Recovery of enzyme activity | |
---|---|
Enzyme | Recovery (%) |
Kallikrein | 95 |
Lipoxidase | 100 |
Catalase | 96 |
Sample load : 0.5mg each
Column : Shodex IEC QA-825 (8.0mmID*75mm) Eluent : (A); 20mM Tris-HCl buffer(pH7.5) (B); (A) + 0.5M NaCl Linear gradient: (A) to (B), 1min