Refractive Index Detector: Shodex RI-201H
A new refractive index (RI) detector series is now available.
(Features)
· uses the same optical system as that of the RI-101.
· Reasonable price
(Specifications)
Product code | ||
---|---|---|
Model |
RI-201H
|
|
Analysis
|
||
Flow cell type |
2 chamber-type
|
|
Measuring method |
Deflection type
|
|
Refractive Index range
|
1.00 to 1.75
|
|
Measuring range |
0.25 to 512 micro-RIU
|
|
Drift |
0.2micro-RIU/h (Pure water 1mL/min, PURGE OFF)
|
|
Linearity range
|
≥600micro-RIU
|
|
Noise |
≤2.5nRIU (Pure water, response : 1.5sec)
|
|
Response |
0.1 , 0.25 , 0.5 , 1 , 1.5 , 2 , 3 , 6 sec
|
|
Auto zero |
Full auto zero
|
|
Auto zero range |
All range
|
|
Off-set range |
10micro-RIU
|
|
Off-set response |
50nRIU
|
|
Integrator output (Sensitivity) |
DC 0 to 1V
(2mV/micro-RIU, 8mV/micro-RIU) |
|
Cell volume |
8micro-L
|
|
Flow rate | (Usual) |
0.2 to 3.0mL/min
|
(Max.) |
10mL/min (solvent: pure water)
|
|
Max. back pressure |
50kPa
|
|
Internal volume |
IN to Cell: ca. 60micro-L
Cell to OUT: ca. 600micro-L All(Cell to OUT: ca.670micro-L |
|
Recorder output |
0 to 10mV/FS
|
|
EXternal input |
–
|
|
EXternal output |
(1) READY (temperature control)
(2) LEAK (3) ERROR (ROM, RAM, PARAMETER, HOME POSITION, OVER-HEAT, OPTI.-BALANCE, INTENSITY) |
|
Temp. control |
OFF, 30 to 50deg-C (1deg-C step), 77deg-C Temp. fuse
|
|
Communication port |
USB
|
|
Operational support functions |
None
|
|
Wetted materials |
Stainless steel 316, Teflon, Quartz glass
|
|
Power source, Power consumption |
AC 100 to 240V±10%, 50/60Hz, 150VA max.
|
|
Demensions, Weight |
W260 x D400 x H150 (mm), ca.12kg
|
|
Accessories |
Power cable, single code, connector tube, fuse, operation manual
|