Effect of Sample Load on Resolution (CM-825)
When IEC CM-825 (a column forweak cation exchange chromatography) is used, the effect of sample load on the resolution is shown. The resolution (Rs) of a mixture of ribonuclease A and (alpha) chmotrypsinogen A remains at approximately 3 until the sample load reaches 1.0mg. Even when the sample load is 5.0mg, the Rs is approximately 1.1, which means the substances can still be separated.
Sample :
Ribonuclease A